PRECESSION ELEPHANT 2 Scan Head 2
3.
NOTICE
Focal point
could cause damage to the sensor of the power measurement device.
•
Place a power measurement device beneath the beam exit such way, that
the laser beam illuminates most part of the sensor.
Do not place the sensor in or near the focal point.
4.
CAUTION
Visible or invisible laser radiation
Damage to eye or skin
•
Avoid eye or skin exposure to direct or scattered radiation.
•
Parametrize the laser for visual set up with average power and pulse
repetition rate as low as possible.
Switch on the laser beam:
In the InScript software in the
Node Properties
view of node
Pens
default
<your laser>
on the
Control
tab, set
Gate control
to
Always on
.
5. Maximize the optical power at the beam exit:
a. Measure the optical power with a power meter device.
Eventually it may be necessary to increase the optical power to get a
significant measurement. After measuring do not forget to reduce the
optical power to the previous value.
b. If your scan head has a motorized attenuator then:
i. In the InScript software in the
Node Properties
view of node
Pens
default
head
on the
A Data
tab, set for
Subframe 3
the
Mode
to
Specified
.
ii. Increase/decrease the
Position
until the optical power is
maximized.
44