Manual, F/T Sensor, Ethernet Axia
Document #9620-05-C-Ethernet Axia-02
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel:+1 919.772.0115 • Fax:+1 919.772.8259 •
C-72
Table 9.3
—moncon.cgi Threshold Conditions
Variable
Name
1
Allowed
Values
Description
Example
mce
n
Integers: 0 or 1
Threshold statement n: enable = 1 or disable = 0
mce0=1
mcx
n
Integers: -1 to 5
Selects the axis evaluated by threshold statement
n
.
mcx0=5
Value
Description
Menu Value
-1
Statement disabled
blank
0
Fx axis
Fx
1
Fy axis
Fy
2
Fz axis
Fz
3
Tx axis
Tx
4
Ty axis
Ty
5
Tz axis
Tz
mcv
n
Integers:
-2147483648 to
+2147483647
Sets the counts value to compare the current axis
value by threshold statement
n
.
mcv0=20
mco
n
Hexadecimal:
0x00 to 0xFF
Sets the output code for threshold statement
n
.
mco0=0x00
Note:
1. where
n
is an integer ranging from 0 to 15 representing the threshold statement index
9.4
CGI Variable: Configurations (config.cgi)
The user can set the calibration and tool transformation for the sensor (refer to
Configurations Page (config.htm)
for related information).
Table 9.4
—config.cgi Variables
Variable Name
Allowed
Values
Description
cfgcalsel
integers: 0 to 1
Sets the calibration used by the sensor.
cfgtdu
integers: 1 to 5
The distance measurement units that are used by the configuration’s
tool transformation.
Value
Description
Menu Value
1
inch
in
2
foot
ft
3
millimeter
mm
4
centimeter
cm
5
meter
m
cfgtau
integers: 1 to 2
The rotation units used by the configuration’s tool transformation.
Value
Description
Menu Value
1
degrees (°)
degrees
2
radians
radians