5.3
Measurement Connectors (Measurement Mode IMPD−EXT)
5
-
6
ZA57630
bias, DC bias setting 0 V, 50 Ω load
Distortion
0.2% or less (no load when 100 kHz or less, BW500 kHz, and
3 Vrms output)
ALC
PORT1 / PORT2 / OFF
The range that can be controlled constantly will be narrower than
the setting range of the measurement signal amplitude depending
on product variations or DUT impedance.
Output limit
Setting range
1 aVrms to 999 GVrms
Setting resolution 3 digits or 1 aVrms, whichever is the largest
Remark
This function prevents an excessive signal level being output due
to an incorrect setting. A measurement signal amplitude exceeding
the output limit setting cannot be output.
・
Normal DC bias
Setting range
−999 GV to +999 GV
However, the setting range is limited to −5.00 × K V to +5.00 × K V
by DUT drive amplifier gain setting K.
There is a following restriction between the measurement signal
amplitude setting and normal DC bias setting.
(Measurement signal amplitude setting [Vrms] × 1.42) + |Normal DC bias
setting [V]| ≤ 5.0×|K|
Setting resolution 3 digits or M
DC
V, whichever is the largest
M
DC
= 10
K"
. Here, K" indicates the maximum integer that does not
exceed log
10
(|K|×10
−2
).
Example: When K = 82, K" = −1 and M
DC
= 100 mV.
Accuracy
±(|1% of normal DC bias setting [V]| + 3% of measurement signal amplitude setting
[Vrms] + 30 mV) When no load
Remark
Either the front panel
H
CUR
/OSC
or rear panel
DC BIAS OUT
is
selected for the output connector of the normal DC bias.
・
Output impedance
50 Ω (nominal value)
・
DUT drive amplifier gain setting K
Set the gain of the amplifier or attenuator that supplies the
measurement signal to the DUT.
The measurement signal amplitude and normal DC bias applied to
the DUT can be set directly.
Setting range
±(1E−12 to 1E+12)
Setting resolution 3 digits or 1E−12, whichever is the largest
Summary of Contents for ZA57630
Page 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Page 2: ......
Page 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Page 22: ...1 8 ZA57630 Blank...
Page 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Page 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Page 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Page 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Page 36: ...2 14 ZA 57630 Blank...
Page 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Page 240: ...XVI ZA57630 Blank...
Page 242: ......
Page 244: ......
Page 245: ......
Page 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...