3.5 Setting Screen Reference
3
-
59
ZA 57630
■ [RESONANT TRACK]
-
[UPPER FREQ]
: Set the tracking frequency upper limit.
■ [RESONANT TRACK]
-
[LOWER FREQ]:
Set the tracking frequency lower limit.
This sets the frequency range for performing resonant frequency tracking.
○ Setting procedure
<Numerical settings (immediate settings)>, <Numerical settings>
○ Functions of function keys
・
[SWEEP -> RANGE]
: Set the same range as the frequency sweep measurement setting.
・
[MKR -> RANGE]
: Set the smaller and larger of MARKER1 and MARKER2 to
LOWER FREQ and UPPER FREQ, respectively.
・
[MKR -> UPPER]
: Set the value of the active marker as the tracking frequency upper limit.
・
[MKR -> LOWER]
: Set the value of the active marker as the tracking frequency lower limit.
・
[EDIT]
: Set the tracking frequency upper and lower limits with the
numerical input palette.
・
[MODIFY]
: Modify the tracking frequency upper and lower limits with the
knob or cursor keys.
■ [RESONANT TRACK]
-
[TARGET PHASE]
: Set the phase for resonant frequency.
This sets the phase for resonant frequency. Usually 0° should be used, but you can also set
the phase shift portion depending on the measurement system.
○ Setting procedure
<Numerical settings (immediate settings)>, <Numerical settings>
○ Functions of function keys
・
[EDIT]
: Set the phase for resonance with the numerical input palette.
・
[MODIFY]
: Modify the phase for resonance with the knob or cursor keys.
The setting range is
–
179.999° to +180.000° and the setting resolution is 0.001°.
■ [RESONANT TRACK]
-
[TOLERANCE]
: Set the permissible error.
This sets the permissible error of the phase for tracking. Measurement is repeated while
changing the frequency until the frequency comes within the permissible error set here.
○ Setting procedure
<Numerical settings (immediate settings)>, <Numerical settings>
The setting range is 0.001° to +179.999° and the setting resolution is 0.001°.
○ Functions of function keys
・
[EDIT]
: Set the permissible error with the numerical input palette.
・
[MODIFY]
: Modify the permissible error with the knob or cursor keys.
Summary of Contents for ZA57630
Page 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Page 2: ......
Page 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Page 22: ...1 8 ZA57630 Blank...
Page 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Page 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Page 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Page 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Page 36: ...2 14 ZA 57630 Blank...
Page 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Page 240: ...XVI ZA57630 Blank...
Page 242: ......
Page 244: ......
Page 245: ......
Page 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...