4.5 Setting Sweep Measurement
4
-
24
ZA57630
When the measurement mode is G−PH, the ratio of V1 and V2 and the phase obtained
through measurement at frequency f [Hz] are R [unitless] and θ[°], respectively. Furthermore,
each parameter is defined as follows.
p =
θ
/ 180 [rad], angular frequency
ω
= 2
f [rad/s]
Table 4-6 Display Parameter Conversion Formulas (Measurement Mode: G
−PH)
Display parameter
Units
Formula
Remarks
R
None
R
Gain
dBR
dB
20×log
10
(R)
Gain
θ
°
θ
Phase
a
None
R cos(p)
Gain (real part)
b
None
R sin(p)
Gain (imaginary part)
GD
s
-
∂p
/
∂ω
Group delay
Summary of Contents for ZA57630
Page 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Page 2: ......
Page 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Page 22: ...1 8 ZA57630 Blank...
Page 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Page 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Page 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Page 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Page 36: ...2 14 ZA 57630 Blank...
Page 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Page 240: ...XVI ZA57630 Blank...
Page 242: ......
Page 244: ......
Page 245: ......
Page 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...