1.3 Operating Principle
1
-
4
ZA57630
1.3 Operating Principle
1.3.1 Operating Principle
The
ZA57630
applies a sine wave signal from an internal measurement signal source to the
device under test (DUT). It then detects current I flowing through the DUT and voltage V
applied to the DUT, performs discrete Fourier transform (DFT), and calculates impedance Z
(= V/I) of each analysis frequency component from the obtained Fourier coefficient obtained
through the DFT. Parameters such as inductance L, capacitance C, and resistance R are
calculated from the impedance (magnitude and phase angle).
Figure 1-1
Block Diagram
H
CUR
OSC
H
POT
PORT1
L
CUR
PORT2
~
o
+
POWER
SUPPLY
DC BIAS
OUT
DFT
OSC
CONTROL
~
LINE
UI CPU
VGA
LCD
PANEL
REMOTE IF
USB
−
B
USB
−
A
GPIB
LAN
RS
−
232
10 MHz
REF IN
10 MHz
REF OUT
MEASUREMENT BLOCK
A/D
A/D
HANDLER
INTERFACE
AUX
L
POT
PORT3
Summary of Contents for ZA57630
Page 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Page 2: ......
Page 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Page 22: ...1 8 ZA57630 Blank...
Page 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Page 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Page 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Page 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Page 36: ...2 14 ZA 57630 Blank...
Page 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Page 240: ...XVI ZA57630 Blank...
Page 242: ......
Page 244: ......
Page 245: ......
Page 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...