3.5 Setting Screen Reference
3
-
92
ZA 57630
■ [LOAD CORR]
-
[LOAD STD FMT]
: Set the load standard value format.
This specifies the format for when the load standard values are input.
○ Setting procedure
<List operations (immediate settings)>
・
Cs-D
: Specify with the series capacitance and dissipation factor.
・
Cp-D
: Specify with the parallel capacitance and dissipation factor.
・
Rp-Cp
: Specify with the parallel resistance and parallel capacitance.
・
Rs-Ls
: Specify with the series resistance and series inductance.
・
Rs-X
: Specify with the resistance (series resistance) and reactance.
・
Z-
θ
: Specify with the impedance and phase.
For details on load correction, refer to
“1.8.5 Load Correction”
in
Instruction Manual (Advanced)
.
■ [LOAD CORR]
-
[LOAD STD VALUE]
: Set the load standard values.
This sets the load standard values.
When the menu is tapped, the
[[LOAD STANDARD VALUE]]
dialog box appears.
Set the frequency and load standard values. Up to 30 sets of load standard values (30
frequencies) can be set.
■ [LOAD CORR]
-
[LOAD STD VALUE]
-
[[LOAD STANDARD VALUE]]
: Specify the load standard values.
Set the frequency and load standard values. Up to 30 sets of load standard values (30
frequencies) can be set.
When the parameter is selected and the knob is turned clockwise, the setting item changes in
the order of No. 1 Frequency → No. 1 Standard Value (Y1) → No. 1 Standard Value (Y2) →
No. 2 Frequency → No. 2 Standard Value (Y1) → No. 2 Standard Value (Y2) → ...
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[DELETE LINE]
: Delete the line.
・
[SORT]
: Sort the displayed information in frequency ascending order.
・
[EDIT]
: Set the load standard values with the numerical input palette.
・
[MODIFY]
: Modify the load standard values with the knob or cursor keys.
The setting range is 10 µHz to 36 MHz and the setting resolution is 10 µHz for the frequency.
The setting range is –1.0 T to +1.0 T and the setting resolution is 6 digits or 1 a (10
–18
),
whichever is larger, for the load standard values.
■ [PORT EXT]
: Set port extension correction.
This sets the port extension correction function. It can be set when the measurement mode is
IMPD−EXT, IMPD−3T and IMPD−2T.
For details on port extension, refer to
“1.8.6 Port Extension”
in
Instruction Manual (Advanced)
.
Summary of Contents for ZA57630
Page 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Page 2: ......
Page 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Page 22: ...1 8 ZA57630 Blank...
Page 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Page 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Page 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Page 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Page 36: ...2 14 ZA 57630 Blank...
Page 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Page 240: ...XVI ZA57630 Blank...
Page 242: ......
Page 244: ......
Page 245: ......
Page 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...