4.2 Simple Operation for First Time Use
4
-
10
ZA57630
procedures, refer to
“1.8 Error Correction”
in
Instruction Manual (Advanced)
.
The included
100 Ω RESISTOR
can be used for checking the operation in the IMPD−3T
measurement mode.
Figure 4-5
How to Check Operation Using 100 Ω RESISTOR (IMPD−3T)
The measurement result of impedance Z and phase θ
Z
are normal if they are roughly in the
following ranges.
1 MHz or less
Z: 100 Ω±2.0 Ω, θ
Z
: 0° ±2°
10 MHz or less Z: 100 Ω±4.0 Ω, θ
Z
: 0°±10°
36 MHz or less Z: 100 Ω±25.0 Ω, θ
Z
: 0°±20°
Measurements at frequencies above 10 MHz are for supplemental value.
H
CUR
/ OSC
H
POT
/ PORT1
L
CUR
/ PORT2
ZA57630
L
POT
100 Ω RESISTOR
(accessory)
Summary of Contents for ZA57630
Page 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Page 2: ......
Page 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Page 22: ...1 8 ZA57630 Blank...
Page 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Page 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Page 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Page 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Page 36: ...2 14 ZA 57630 Blank...
Page 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Page 240: ...XVI ZA57630 Blank...
Page 242: ......
Page 244: ......
Page 245: ......
Page 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...