4.7 Equalization
FRA5087
4-22
4.7.2 Principle of equalization
The equalization principle is described below with respect to Fig. 4-14.
OSC
CH1
CH2
FRA5087
Amp
Probe1
Probe2
Fdut
Vin
Vout
V1m
V2m
OSC
CH1
CH2
FRA5087
Amp
Probe1
Probe2
Veql
V1e
V2e
a) Measurement of DUT b) Measurement of measurement system
Fig. 4-14 Principle of equalization
1)
Measure by using the connection shown in Fig. 4-14b. Set the Amp output voltage level
at Veql to have CH1 and CH2 input voltage levels, V1e and V2e, respectively, to be the
following:
V1e = Veql
Probe1
V2e = Veql
Probe2
2)
Record/store the measured data above in the EQL memory as the equalization data. The
data for CH2/CH1 are recorded/stored in the EQL memory. Therefore, the content of the
EQL memory will be as follows:
EQL
V2e
V1e
Probe2
Probe1
3)
Connect as shown in Fig. 4-14a and measure the overall dut system. Putting the Amp
output voltage to be Vin and the Fdut output voltage to be Vout, you can obtain input
voltages of CH1 and CH2, V1m and V2m, respectively, as follows:
V1m = Vin
Probe1
V2m = Vout
Probe2 = Vin
Fdut
Probe2
4)
Equalize these data by using the data stored in the EQL memory. Since the actual
processing of equalization comprises division operations, i.e., the CH2 measurement value
divided by the value stored in the EQL memory, CH1 and CH2 voltage levels after
equalization, V1 and V2, respectively, will be as follows:
V1 V1m Vin Probe1
V2 =
V2m
EQL
Vin Fdut Probe2
Probe2 / Probe1
Vin Fdut Probe1
5)
Display the ratio of CH1 to CH2 (i.e., ratio of V1 to V2) to cancel the effects of Vin and
Probe1. The displayed data show Fdut, the characteristics of the DUT (device under test).
Note, however, that the effect of Probe1 will be left unremoved, if the absolute value of
CH1 or CH2 is displayed.
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......