9.4 Analysis process
FRA5087
9-10
9.4 Analysis process
●Analysis mode
Ratio
・・・・・・
CH1/CH2, CH2/CH1
Level
・・・
・・
CH1, CH2
●Measurement error
When a calibration is executed shortly prior, and when the channels are either CH1/CH2 or
CH2/CH1, and when the analysis input voltage is between 100 mVpeak and 10 Vpeak (no more
than 2 Vpeak if 2.2 MHz is being exceeded):
≦
20 kHz
≦
500 kHz
≦
2.2 MHz
>2.2 MHz
A, b, R
±0.5 %
±1 %
±10 %
±25 %
dBR
±0.05 dB
±0.1 dB
±1 dB
±2 dB
Phase
±0.3 °
±0.5 °
±2 °
±5 °
9.5 Calculations
●Four arithmetic operations
Four arithmetic operations between: data and data, data and numeric, numeric and numeric
●Derivation and integration
Data derivation, second order differential, integration, second order integration
●Open loop and closed loop conversion
Open loop -> closed loop conversion, closed loop -> open loop conversion
Feedback characteristics is specified by data or numeric
9.6 Auto sequence
●Key sequence record
Records the key operations in the internal memory
●Key sequence execution
Executes previously recorded key operations
●Key sequence deletion
Deletes recorded key operations
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......