5.4 Operation method
FRA5087
5-9
5.4.3 Graph screen description
5.4.4 Shunt resistance current-voltage conversion coefficient setting
Use the menu [INPUT] [WEIGHTING FACTOR] to set current-voltage conversion coefficient.
Setting range is 0 to 1.0E+6.
Voltage input is CH1. If connecting a preamplifier with a gain of 100, set the CH1 weighting
factor to 0.01 (inverse of 100).
Current input is CH2. If the CH2 shunt resistance is 100 m
Ω
, set the CH2 weighting factor to
10 (inverse of 100 m).
Set invert on to invert the phase (+180
). This can be effective for inverting voltage and current
phase when measuring impedance.
Fig. 5-1 Phase invert connection
Impedance units
Graph and impedance units are shown
Vertical axis scale can also be shown with log spacing
OSC: 100.0000000MHz 7.00 Vpeak DC 1.00 V INTEG: 100cycle HMNC: 1
SWP: 100steps/sweep CPRSN:OFF SLSWP:OFF ANAL:CH2/CH1 EQL:OFF
SWEEP
STOP
1
2
3
4
5
6
PHASE[deg]
FREQUENCY(Hz)
IMPD[
¶
n
1m
10m
100m
10
100
1k
100k
-90
-75
-60
-45
-30
-10
0
*f: 1.0000000kHz
@
@
@*R: 7.5 m
¶
@
@*
Æ : -80.66 deg
INVERT:OFF
CH1(V)
DUT
SHUNT
CH2(I)
+
+
-
-
d
³
Æ
d
¬
Ì
ü
Í
O
h
¤
Ê
INVERT:ON
CH1(V)
DUT
SHUNT
CH2(I)
+
-
-
+
Common ground for voltage
and current inputs
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......