9.3 Measurement process
FRA5087
9-7
9.3 Measurement process
●Mode
REPEAT
Repeatedly measures at a fixed frequency
SINGLE
Measures once at a fixed frequency
SWEEP
Measures while sweeping between the upper limit frequency and
the lower limit frequency
●Auto ranging feature
A feature that switches the input range according to the input signal levels
●Delay feature
A feature that inserts a delay after changing the frequency before starting the measurement
Delay quantity is set either by the length of the time or by the number of cycles
During frequency sweeping, the following sequence is repeated: frequency measurement ->
delay -> measurement.
Second setting
Setting range
0 to 9,999 seconds
Setting resolution
10ms
Cycle setting
Setting range
0 to 9,999 cycles
Setting resolution
1 cycle
●Integration feature
A data integration feature for measuring while eliminating noise effects
Measurement repetition is set either by the length of the time or by the number of cycles.
Cycle setting
Setting range
0 to 9,999 cycles
Setting resolution
1 cycle
Second setting
Setting range
0 to 9,999 seconds (one cycle of integration is always executed
regardless of the setting)
Setting resolution
10 ms
●Automatic integration feature
A feature that repeats integration until reaching a certain reliability
Maximum integration time
Cycle setting
Setting range
2 to 9,999 cycles
Setting resolution
1 cycle
Second setting
Setting range
0 to 9,999 seconds (two cycles of integration are always
executed regardless of the setting)
Setting resolution
10 ms
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......