8.5 Performance test
FRA5087
8-7
8.5.5 Oscillator distortion
Oscillator sine wave distortion is tested.
a)
Setting
FRA5087
Distortion meter
Output waveform
Sinusoidal wave
Noise distortion (THD) measuring mode
Output voltage
AC
10 Vpeak
DC bias
0 V
b)
Connection
FRA5087
Distortion meter
BNC-BNC
Coaxial cable
OSC
CH1 CH2
c)
Procedure
Set the oscillator frequency as follows and read the distortion meter indication. Set the
distortion meter lowpass filter (LPF) to 100 kHz.
Oscillator frequency: 10 kHz
d)
Judgment
Oscillator frequency
Distortion (THD)
Tolerance
10 kHz
%
0.2 %
THD: Total harmonic distortion