8.5 Performance test
FRA5087
8-9
8.5.7 Analyzer IMRR
IMRR at 60 Hz is tested.
a)
Setting
FRA5087 oscillator
FRA5087 analyzer
Frequency
60 Hz
Integration
100 cycles
Output waveform Sinewave
Measuring mode
CH1, CH2
Output voltage
Display mode
X axis : logF
AC
10 Vpeak
Y1 axis : dBR
DC bias
0 V
Y2 axis :
b)
Connection
FRA5087
BNC-BNC
coaxial cable
BNC T-type
divider
BNC-alligator clip
coaxial cable
BNC shorting
plug
OSC
CH1 CH2
FRA5087
OSC
CH1 CH2
CH1 IMRR measurement CH2 IMRR measurement
c)
Procedure
Set the analyzer mode as follows. Conduct Single measurement with FRA5087.
Read the dBR measuring results from the LCD screen.
Analyzer mode for CH1 IMRR measurement: CH2/CH1
Analyzer mode for CH2 IMRR measurement: CH1/CH2
d)
Judgment
Connection
Measured value
Tolerance
CH1 (analyzer mode: CH2/CH1)
dBR= dB
More than 120 dB
CH2 (analyzer mode: CH1/CH2)
dBR= dB
More than 120 dB
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......