8.5 Performance test
FRA5087
8-10
8.5.8 Analyzer dynamic range
The dynamic range is tested.
a)
Setting
FRA5087 oscillator
FRA5087 analyzer
Sweep upper limit frequency
10 MHz
Integration
2,000 cycles
Sweep lower limit frequency
10 Hz
Measuring mode
CH1, CH2
Output waveform
Sinewave
Display mode
X axis :
logF
Output voltage
Y1 axis: dBR
AC
10 Vpeak
DC bias
0 V
b)
Connection
FRA5087
BNC-BNC
coaxial cable
BNC
shorting plug
OSC
CH1 CH2
FRA5087
OSC
CH1 CH2
CH1 dynamic range measurement
CH2 dynamic range measurement
c)
Procedure
Set the analyzer mode as follows and sweep with the FRA5087.
Read the minimum dBR measuring results from the LCD screen.
Analyzer mode for measuring CH1 dynamic range: CH2/CH1
Analyzer mode for measuring CH2 dynamic range: CH1/CH2
d)
Judgment
Connection
Minimum measured value
Tolerance
CH1 (10 Hz to 1 MHz)
dBR= dB
140 dB typ
CH2 (10 Hz to 1 MHz)
dBR= dB
140 dB typ
CH1 (1 MHz to 10 MHz)
dBR= dB
80 dB typ
CH2 (1 MHz to 10 MHz)
dBR= dB
80 dB typ
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......