8.5 Performance test
FRA5087
8-6
8.5.4 Oscillator output amplitude accuracy
The oscillator output amplitude and frequency response are tested.
a)
Setting
FRA5087
Multi-meter
Output waveform sinusoidal wave
Measurement mode
AC voltage
Output voltage
(True
RMS)
AC
10 Vpeak
DC bias
0 V
b)
Connection
FRA5087
Multi-meter
BNC-BNC
Coaxial cable
OSC
CH1 CH2
c)
Procedure
Set the oscillator frequency as follows and read the multi-meter indications. Use a
wideband multi-meter for frequencies above 100 kHz.
Oscillator frequency: 1 kHz, 100 kHz, 1 MHz, 10 MHz
Derive the amplitude from the multi-meter indication by the following formula.
[Amplitude accuracy (dB) = 20
×
log10[multi-meter indication (Vrms)] - 16.9897
d)
Judgment
Oscillator frequency
Multi-meter
Amplitude accuracy
Tolerance range
1 kHz
Vrms
dB
-0.3 – +0.30 dB
100 kHz
Vrms
dB
-0.3 – +0.30 dB
1 MHz
Vrms
dB
-1.00 – +1.00 dB
10 MHz
Vrms
dB
-3.00 – +3.00 dB
Summary of Contents for FRA5087
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 2: ......
Page 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Page 40: ...2 6 Calibration FRA5087 2 12 Blank...
Page 120: ...4 19 Condition display FRA5087 4 46 Blank...
Page 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Page 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Page 170: ...8 5 Performance test FRA5087 8 12 Blank...
Page 190: ......
Page 192: ......
Page 193: ......