3.6 Basic Operation Examples
CA5350
3-21
Set appropriate value according to the properties of the signal.
Any random filter setting can be made irrespective of
I/V GAIN
settings. However,
overall response time of
CA5350
is the combination of the response of
I/V GAIN
and the
response time fo filter. Generally, it will be decided by the one with longer response
time.
Setting to Auto Filter
Selecting
AUTO
in the Menu
FILTER
–ON
,
AUTO
will make filter settings corresponding
to
I/V GAIN
settings irrespective of the filter rise time set in the
MANUAL
mode.
Table 3-2 I/V Gain Settings and Auto-Filter Settings
I/V Gain Settings
(V/A)
Filter Settings
10G (10
10
)
100µs
1G (10
9
)
100µs
100M (10
8
)
30µs
10M (10
7
)
10µs
1M (10
6
)
10µs
100k (10
5
)
3µs
10k (10
4
)
1µs
Setting to auto filter will automatically change the filter settings when
I/V GAIN
settings are changed. Output amplifire gain setting is irrelevant.