3.6 Basic Operation Examples
CA5350
3-17
3.6.2 Simplified Operating Method When You Use Device for the First Time
This section describes the simplified operating method when you use the
CA5350
for the
first time. Do not connect the remote control interface.
Operate as the following steps:
1) Initialize
2) Connected to the sensor (signal source)
3) Zero check OFF
The first step can be passed if the product is in the factory default setting.
Initialization
Turn on the power supply of
CA5350
. After initialization has completed, select the
setting memory No. 0 from the menu
CONFIGURATION
–RECALL
.
C O N F I G U R A T I O N
S
R E C A L L
S
0
Reset to the factory default settings by pressing the
ENTER
key.
Connection with a sensor (or source of signal)
Connect the input connector
(
CURRENT
INPUT
)
in the front panel of
CA5350
and sensor
using a coaxial cable. In the factory default setting of
CA5350
, input is selected to
CURRENT
INPUT
connector located on the front panel.
Similarly, connect the output connector
(
INVERTING
OUTPUT
)
of
CA5350
and
measurement equipment using a coaxial cable. Either front or rear output connector can
be used.
Because
ZERO CHECK
of
CA5350
is still ON, signal from sensor has not been input to
CA5350
. Therefore, output voltage is almost 0V regardless of the sensor current.
Turning Zero Check to OFF
Select
OFF
from the Menu
ZERO
CHECK
.
Z E R O
a
C H E C K
* O N
S
O F F
Press the
ENTER
key to turn
OFF ZERO CHECK
.
A voltage signal converted from the input current signal will be output to the output
connector (
INVERTING
OUTPUT
) of
CA5350
.