7.5 Performance Testing
CA5350
7-5
7.5.1 Current suppression setting accuracy, gain accuracy
In this section, easy to measures the total accuracy of the setting accuracy of current
suppression and gain accuracy of amplifier. For the exact test, please ask a test to the
NF Corporation.
Connection:
Connect
INVERTING
OUTPUT
(Front) to DC voltmeter to BNC cable
Setting:
FILTER
:
MANUAL
,
1ms
Measurement: Measure the
INVERTING
OUTPUT
voltage with the DC voltmeter.
Evaluation:
If the values indicated on DC voltmeter are within the specification
range in table, it is normal.
Current suppression
Amplifier gain
Values indicated on
DC voltmeter
Specification
Range
Setting
Gain
(V/A)
Output
amplifier
gain
8nA
+0.900nA
10G
×1
+8.490V ~ +9.510V
±(5.33%+30mV)
8nA
+0.000nA
10G
×1
-0.150V ~ +0.150V
±150mV
8nA
+8.000nA
1G
×1
+7.648V ~ +8.352V
±(4.15%+20mV)
8nA
+0.000nA
1G
×1
-0.032V ~ +0.032V
±32mV
80nA
+80.00nA 100M
×1
+7.808V ~ +8.192V
±(2.15%+20mV)
80nA
+0.00nA
100M
×1
-0.032V ~ +0.032V
±32mV
800nA
+800.0nA
10M
×1
+7.880V ~ +8.120V
±(1.25%+20mV)
800nA
+0.0nA
10M
×1
-0.032V ~ +0.032V
±32mV
8µA
+8.000µA
1M
×1
+7.900V ~ +8.100V
±(1.0%+20mV)
8µA
+0.000µA
1M
×1
-0.032V ~ +0.032V
±32mV
80µA
+80.00µA
100k
×1
+7.900V ~ +8.100V
±(1.0%+20mV)
80µA
+0.00µA
100k
×1
-0.032V ~ +0.032V
±32mV
800µA
+800.0µA
10k
×1
+7.900V ~ +8.100V
±(1.0%+20mV)
800µA
+0.0µA
10k
×1
-0.032V ~ +0.032V
±32mV
80µA
+80.00µA
10k
×10
+7.180V ~ +8.820V
±(7.75%+200mV)