1.4 Principle of Operation
CA5350
1-5
FILTER
It is equipped with filter that allows setting the rise time in the range of 1µs ~ 300ms in
by 1-3 sequence. You can remove unnecessary noise components and extract the
required signal component.
BIAS
For sensors those require DC bias, voltage bias source equipped that allows to output
any DC voltage in the range of -8V to +8V (resolution 0.001V) will be useful.
SYSTEM CONTROLLER
System controller sets and controls the analog parts according to user operations, and it
displays various information on the front panel LCD. Besides, it can also communicate
(remote I/F control) with the host controller.