5.6 Description of Individual Commands
CA5350
5-17
U
– Queries of Current Suppression Setting
Description: Send the current suppression setting value to the controller. Even if
current suppression is OFF, current suppression value set using S
command will be output.
Response message:
N
D C I
+
0 . 0 0 0
E - 0 3 <term>
Exponential part and mantissa of response message show the following according to the
current suppression range.
Range
Mantissa
Exponential Part
±
8nA range
0.000
E-09
±
80nA range
00.00
E-09
±
800nA range
000.0
E-09
±
8µA range
0.000
E-06
±
80µA range
00.00
E-06
±
800µA range
000.0
E-06
Remarks:
For message terminator refer to
Y
command, and for EOI refer to
K
command.
Message Terminator,
EOI
Exponential Part Engineering Notation
Mantissa (significant 4 digit)
Sign "
+
" or "
-
"
"
DCI
" Fixed
定
Status "
N
": Normal, "
O
": Over