5.6 Description of Individual Commands
CA5350
5-15
N
– Current Suppression ON/OFF
Description: ON, OFF setting of current suppression
Setting:
N0
Current suppression OFF (factory default setting)
N1
Current suppression ON
N2
Auto current suppression starts
Remarks:
Current value when turned on is set with
S
command.
When performing auto current suppression
(N2)
, Zero Check must be Off
(C0)
. For the details of the current suppression, refer to
“4.2 Cancelling
the dark current of sensor”
.
P
– Filter ON/OFF
Description: ON, OFF setting of filter
Setting:
P0
Filter OFF (factory default setting)
P1
Filter ON
Remarks:
Set filter rise time using
T
Command with filter ON.
R
– I/V Gain Setting
Description: Setting of I/V gain
Setting:
R4
10k (=10
4
) V/A (factory default setting)
R5
100k (=10
5
) V/A
R6
1M (=10
6
) V/A
R7
10M (=10
7
) V/A
R8
100M (=10
8
) V/A
R9
1G (=10
9
) V/A
R10
10G (=10
10
) V/A
Remarks:
Gains can be set as above only when the output amplifier gain is
×
1.
When output amplifier gain is set to ×10, total gain will be 10 times of
the value set by this command.
Settings except the above will cause error.