7.5 Performance Testing
CA5350
7-6
7.5.2 Output offset voltage
Connection:
Connect
INVERTING
OUTPUT
(Front) to DC voltmeter to BNC cable
Setting:
FILTER
ON
,
MANUAL
Measurement: Measure the
INVERTING
OUTPUT
voltage with the DC voltmeter.
Evaluation:
If the values indicated on DC voltmeter are within the specification
range in table, it is normal.
Filter setting
Values indicated on
DC voltmeter
Specification
30µs
-20mV ~ +20mV
±20mV
3ms
10ms
7.5.3 DC bias voltage setting accuracy
Connection:
Connect
INVERTING
BIAS
OUTPUT
(Front) to DC voltmeter to BNC
cable
Setting:
BIAS
:
ON
Measurement: Measure the
INVERTING
BIAS
OUTPUT
voltage with the DC voltmeter.
Evaluation:
If the values indicated on DC voltmeter are within the specification
range in table, it is normal.
DC bias voltage setting
Values indicated on
DC voltmeter
Specification
-8.000V
+7.900V ~ +8.100V
±(1%+20mV)
0.000V
-20mV ~ +20mV
±20mV
+8.000V
-8.100V ~ -7.900V
±(1%+20mV)
Note: Voltage of reverse polarity of the setting will appear.