National Instruments Corporation
37
NI 5450 Calibration Procedure
10. Configure the horizontal properties using the niScope Configure
Horizontal Timing VI.
11. Set Time Interleaved Sampling using the niScope Property Node.
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
HorizontalTiming
with the following
parameters:
vi
: The instrument handle from
niScope_init
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
minSampleRate
: The
Sample Rate
value
listed in Table 12 (NI 5152), Table 13
(NI 5153), or Table 14 (NI 5154) for the
current iteration
refPosition
:
50.0
minNumPts
:
300,000
LabVIEW VI
C/C++ Function Call
Call
niScope_SetAttribute
ViBoolean
with the following
parameters:
vi
: The instrument handle from
niScope_init
channelList
:
"0"
attributeID
:
NISCOPE_ATTR_
ENABLE_TIME_INTERLEAVED_
SAMPLING
value
: The
TIS Enabled
value listed
in Table 12 (NI 5152), Table 13
(NI 5153), or Table 14 (NI 5154) for
the current iteration