NI 5450 Calibration Procedure
3.
Configure the common vertical properties using the niScope Configure
Vertical VI.
4.
Configure the horizontal properties using the niScope Configure
Horizontal Timing VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVertical
with the following parameters:
vi
: The instrument handle from
niScope_init
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
channelList
: The
ChannelList
value listed
in Table 15 (NI 5152) or Table 16
(NI 5153/5154) for the current iteration
range
: The
Range
value listed in Table 15
(NI 5152) or Table 16 (NI 5153/5154) for
the current iteration
offset
:
0.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
NI-SCOPE Function Call
Call
niScope_Configure
HorizontalTiming
with the following
parameters:
vi
: The instrument handle from
niScope_init
enforceRealtime
:
NISCOPE_VAL_FALSE
numRecords
:
1
minSampleRate
:
20,000,000,000
refPosition
:
0.0
minNumPts
:
2,000