NI 5450 Calibration Procedure
Complete Calibration
Performing a complete calibration is the recommended way to warrant that
the NI 5152/5153/5154 will meet or exceed its published specifications for
a two-year calibration interval. At the end of the complete calibration
procedure, you verify that the measurement error falls within the
calibration test limits. Figure 1 shows the programming flow for complete
calibration.
Figure 1.
Complete Calibration Programming Flow
Document
Post-Adjustment
Results
Document
Pre-Adjustment
Results
Calibration/
Verification
Complete
Yes
No
Self-Calibrate
Verify
Meets
Calibration
Test Limits?
Review Verification/
Adjustment Procedure
or Return Device
Adjust (Calibration Dates
and Temperatures Updated)
Verify