NI 5450 Calibration Procedure
56
ni.com
3.
Configure the common vertical properties using the niScope Configure
Vertical VI.
4.
Configure the horizontal properties using the niScope Configure
Horizontal Timing VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVertical
with the following parameters:
vi
: The instrument handle from
niScope_init
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
channelList
: The
channelList
value from
Table 17 for the current iteration.
range
:
1
offset
:
0.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
HorizontalTiming
with the following
parameters:
vi
: The instrument handle from
niScope_init
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
minSampleRate
:
1,000,000,000
refPosition
:
50.0
minNumPts
:
1,000