NI 5450 Calibration Procedure
34
ni.com
Bandwidth
Complete the following steps to verify the bandwidth of the
NI 5152/5153/5154. You must verify both channels with each iteration
listed in Table 12 (NI 5152), Table 13 (NI 5153), or Table 14 (NI 5154).
1.
Open a session and obtain a session handle using the niScope
Initialize VI.
2.
Configure the input impedance and the maximum input frequency
using the niScope Configure Chan Characteristics VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the following
parameters:
resourceName
: The device name assigned
by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the following
parameters:
vi
: The instrument handle from
niScope_init
channelList
:
"0"
inputImpedance
: The
Input Impedance
value from Table 12 (NI 5152), Table 13
(NI 5153), or Table 14 (NI 5154) for the
current iteration
maxInputFrequency
: The
Max Input
Frequency
value listed in Table 12
(NI 5152), Table 13 (NI 5153), or
Table 14 (NI 5154) for the current
iteration