NI 5450 Calibration Procedure
52
ni.com
Compare the resulting percent to the
Calibration Test Limits
or the
Published Specifications
listed in Table 15 (NI 5152) or Table 16
(NI 5154). If the result is within the selected test limit, the device has
passed this portion of the verification.
20. Repeat steps 2 through 19, for each iteration in Table 15 (NI 5152) or
Table 16 (NI 5153/5154).
21. End the session using the niScope Close VI.
You have finished verifying the trigger accuracy for the
NI 5152/5153/5154.
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the following
parameters:
vi
: The instrument handle from
niScope_init