NI 5450 Calibration Procedure
6
ni.com
Table 2.
Required Equipment Specifications for NI 5152/5153/5154 External Calibration
Required
Equipment
Recommended
Equipment
Parameter
Measured
Specification
Signal
Generator
NI 5152/5153
:
Fluke 9500B
oscilloscope
calibrator
or
Wavetek 9500
with
high-stability
reference
option
Fluke 9510
Test Head
NI 5154
:
Fluke
9500B/1100
oscilloscope
calibrator
or
Wavetek
9500/1100 with
high-stability
reference option
Fluke 9510 Test
Head
DC
Accuracy
DC ±(0.025% + 25
μ
V)
into 1 M
Ω
or 50
Ω
Bandwidth,
Trigger
Sensitivity
NI 5152
:
±
2%
output amplitude
flatness for leveled sine
wave up to 300 MHz
relative to 50 kHz into
50
Ω
NI 5153
:
±
3%
output amplitude
flatness for leveled sine
wave up to 500 MHz
relative to 50 kHz into
50
Ω
NI 5154
:
±
4%
output amplitude
flatness for leveled sine
wave up to 1100 MHz
relative to 50 kHz into
50
Ω
Timing
±2 ppm frequency
accuracy
NI 5402 Function Generator
or
Agilent 33220A Function Generator
Trigger
Accuracy
±5% output amplitude
flatness for leveled sine
wave up to 10 V
pk-pk
and
11 MHz relative to
50 kHz into 50
Ω
(3) BNC Cables
—
—
50
Ω
, identical in length
and cable material