NI 5450 Calibration Procedure
You have finished verifying the vertical gain accuracy of the
NI 5152/5153/5154.
Programmable Vertical Offset Accuracy
(NI 5152 Only)
Complete the following steps to verify the programmable vertical offset
accuracy for each NI 5152 channel.
1.
Open a session and obtain a session handle using the niScope
Initialize VI.
2.
Configure the input impedance and the maximum input frequency
using the niScope Configure Chan Characteristics VI.
2
0.9
–0.9
1.732%
–1.732%
2.900%
–2.900%
5
2.25
–2.25
1.732%
–1.732%
2.900%
–2.900%
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the following
parameters:
resourceName
: The device name assigned by
MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the following
parameters:
vi
: The instrument handle from
niScope_init
channelList
:
"0"
inputImpedance
:
NISCOPE_VAL_1_MEG_OHM
maxInputFrequency
:
20,000,000
Table 8.
NI 5153/5154 Vertical Gain Stimuli, Calibration Test Limits, and Published Specifications (Continued)
Range
(V
pp
)
Positive
Input
(V)
Negative
Input
(V)
Calibration Test Limits
Published Specifications
Positive
Negative
Positive
Negative