NI 5450 Calibration Procedure
22
ni.com
16. Move the calibrator test head to the digitizer input channel 1 and repeat
steps 2 through 15 for every configuration in Table 3 (NI 5152) or
Table 4 (NI 5153/5154), changing the value of the
channelList
parameter from
"0"
to
"1"
.
17. End the session using the niScope Close VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the following
parameter:
vi
: The instrument handle from
niScope_init
Table 7.
NI 5152 Vertical Gain Stimuli, Calibration Test Limits, and Published Specifications
Range
(V)
Positive
Input
(V)
Negative
Input
(V)
Calibration Test Limits
Published Specifications
Positive
Negative
Positive
Negative
0.1
0.045
–0.045
1.00%
–1.00%
1.26%
–1.26%
0.2
0.09
–0.09
1.00%
–1.00%
1.26%
–1.26%
0.4
0.18
–0.18
1.00%
–1.00%
1.26%
–1.26%
1
0.45
–0.45
1.00%
–1.00%
1.26%
–1.26%
2
0.9
–0.9
1.00%
–1.00%
1.26%
–1.26%
4
1.8
–1.8
1.00%
–1.00%
1.26%
–1.26%
10
4.5
–4.5
1.00%
–1.00%
1.26%
–1.26%
Table 8.
NI 5153/5154 Vertical Gain Stimuli, Calibration Test Limits, and Published Specifications
Range
(V
pp
)
Positive
Input
(V)
Negative
Input
(V)
Calibration Test Limits
Published Specifications
Positive
Negative
Positive
Negative
0.1
0.045
–0.045
1.025%
–1.025%
2.200%
–2.200%
0.2
0.09
–0.09
1.025%
–1.025%
2.200%
–2.200%
0.5
0.225
–0.225
1.025%
–1.025%
2.200%
–2.200%
1
0.45
–0.45
1.025%
–1.025%
2.200%
–2.200%