NI 5450 Calibration Procedure
28
ni.com
Timing Accuracy
Complete the following steps to verify the timing accuracy for the
NI 5152/5153/5154.
1.
Open a session and obtain a session handle using the niScope
Initialize VI.
Table 9.
NI 5152 Programmable Vertical Offset Accuracy Limits
Iteration
Range
(V
pp
)
Positive
Offset
(V)
Negative
Offset
(V)
Calibration Test
Limits
Published
Specifications
Positive
Negative
Positive
Negative
1
0.1
0.9
–0.9
0.8%
–0.8%
0.9%
–0.9%
2
0.2
0.9
–0.9
0.8%
–0.8%
0.9%
–0.9%
3
0.4
0.9
–0.9
0.8%
–0.8%
0.9%
–0.9%
4
1
0.9
–0.9
0.8%
–0.8%
0.9%
–0.9%
5
2
9
–9
0.8%
–0.8%
0.9%
–0.9%
6
4
9
–9
0.8%
–0.8%
0.9%
–0.9%
7
10
9
–9
0.8%
–0.8%
0.9%
–0.9%
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the following
parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE