NI 5450 Calibration Procedure
3.
Configure the common vertical properties using the niScope Configure
Vertical VI.
4.
Configure the horizontal properties using the niScope Configure
Horizontal Timing VI.
5.
Commit all the parameter settings to hardware using the niScope
Commit VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVertical
with the following parameters:
vi
: The instrument handle from
niScope_init
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
channelList
:
"0"
range
: The
Range
value listed in Table 12
(NI 5152), Table 13 (NI 5153), or Table 14
(NI 5154) for the current iteration
offset
:
0.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
HorizontalTiming
with the following
parameters:
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
vi
: The instrument handle from
niScope_init
minSampleRate
:
10,000,000
refPosition
:
50.0
minNumPts
:
30,000
LabVIEW VI
C/C++ Function Call
Call
niScope_Commit
with the
following parameter:
vi
: The instrument handle from
niScope_init