40
3.
Orient the stack tube so that the chamfers of the plates are toward the back of the instrument and the bar code labels are to
the left.
NOTE: A label on the inside of the stack tube reminds users of the correct orientation for loading plates. The stack tube itself
will only fit on the interface plate in one orientation.
4.
Slide the stack tube carefully over the top of the stack of plates such that the plates slide into the tube. Once the stack tube
base is flush with the stable surface, all but the extra plate will be loaded into the stack tube (
5.
Remove the single plate adaptor if it is present on the stacker.
6.
Place the full stack tube on the input (right) stacker interface plate.
7.
Place an empty stack tube of the same size on the output (left) stacker interface plate.
Figure 6:1 Loading MULTI-ARRAY plates into stack tube
If the stack of plates to be run is too tall for the above procedure, you can insert the additional plates through
the top of the stack tube (up to the stack tube capacity) during the run. Be careful to not to add more plates than
can be accommodated in the output stack tube.
CAUTION:
When reading multiple plates in a run, ensure that the stack tube is not overloaded. See
WARNING:
Loading a tall stack of plates could lead to spilling potentially harmful chemical reagents. Use caution when
loading the stack tubes.
Summary of Contents for SECTOR Image 2400
Page 1: ...SECTOR Imager Models2400 6000 INSTRUMENT MANUAL www mesoscale com ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 ImportantInformation ...
Page 15: ...15 3 SystemDescription ...
Page 28: ...28 4 Installation ...
Page 30: ...30 5 QuickStart ...
Page 36: ...Quick Start 36 Figure 5 8 SECTOR Imager window Run Options dialog box ...
Page 38: ...38 6 UsingtheSECTORImager ...
Page 42: ...42 7 Maintenance ...
Page 45: ...45 8 Appendix ...
Page 54: ...54 9 TechnicalSupport ...
Page 56: ...Technical Support 9 2 ...