Important Information
14
CAUTION:
To avoid interference from voltage transients, the computer and instrument should be powered by the same
electrical circuit. This may be accomplished using an uninterruptible power supply (UPS), power strip of proper electrical rating,
or by verifying the power wall outlets. If one of these is not possible, a ground strap should be used to connect the instrument
chassis to the computer chassis. Please contact MSD
Scientific and Technical Support
if you have questions or need
assistance.
2.3
Software Compatibility
The Microsoft Windows
®
Update service can affect the operation of the instrument software. When running, the update application can
turn off the database services upon which the instrument relies, causing errors and stopping instrument operation. Updates should not
be run during instrument use. Run manually or schedule Windows updates when you are sure that the instrument will not be in use.
Occasionally, Windows Update does not restart the database server after it runs. This will cause the instrument software to run
incorrectly. Restarting the computer will restart the database service and restore the instrument to full operation.
The SECTOR Imager system has not been tested for compatibility with all programs. Installation of additional applications, such as
anti-virus and security programs may interfere with function.
2.4
Decontamination Prior to Shipping or Servicing
The SECTOR Imager may have been used to analyze infectious materials or in an environment where infectious materials were
handled. For the protection of future users and service personnel, please follow site safety procedures and the directions of the site
safety officer to disinfect the SECTOR Imager before shipping or service. Please contact MSD
WARNING:
Follow site safety procedures and the directions of the site safety officer to determine decontamination
requirements for the SECTOR Imager before shipping or service.
Summary of Contents for SECTOR Image 2400
Page 1: ...SECTOR Imager Models2400 6000 INSTRUMENT MANUAL www mesoscale com ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 ImportantInformation ...
Page 15: ...15 3 SystemDescription ...
Page 28: ...28 4 Installation ...
Page 30: ...30 5 QuickStart ...
Page 36: ...Quick Start 36 Figure 5 8 SECTOR Imager window Run Options dialog box ...
Page 38: ...38 6 UsingtheSECTORImager ...
Page 42: ...42 7 Maintenance ...
Page 45: ...45 8 Appendix ...
Page 54: ...54 9 TechnicalSupport ...
Page 56: ...Technical Support 9 2 ...