Important Information
10
Important Information
2
2.1
General Operation
The SECTOR Imagers are designed to run 24 hours a day. Once initialized, the SECTOR Imager enters standby mode when plates are
not running. Standby maintains the CCD detector temperature. The SECTOR Imager requires no additional initialization when left in
standby mode.
WARNING:
The instrument can only be operated with all covers in place. If the unit is operated in any manner not specified
in this manual, the protection provided by the equipment may be impaired.
CAUTION:
Installation of additional software on the computer system used to operate the SECTOR Imager is not supported.
Specifically, updating aspects of the operating system or installing any software that changes parameters of the computer
environment could interfere with proper operation of the instrument software.
CAUTION:
Running screen-savers, automated maintenance software, network-security software, and other possible
software on the SECTOR Imager computer system could cause conflicts with the operation of the instrument software.
CAUTION:
The instrument generates files when running. Virus scanning on the instrument computer can interfere with
instrument operation. It is suggested that any virus checking software run on the instrument be configured to minimize the
scanning of newly generated files while the instrument is operating.
CAUTION:
Use of other applications while plates are being read may interfere with system performance. Use of operating
system power features that disable USB communication, such as Hibernate, will cause the system to stop responding and
requires a power off/on cycle of the instrument and PC to restore function.
CAUTION:
Additional USB devices may
not
be connected to the SECTOR Imager computer system.
CAUTION:
Changes to the computer clock can cause a system error if the changes are made during a plate read. A system
error may also occur if a plate is being read when the time is automatically changed from standard to daylight savings time or
vice versa.
NOTE: The SECTOR Imager should be operated in a dust-free environment with an ambient temperature between 20–26
o
C, and
humidity levels between 10–80% (non-condensing). Environments or locations with high levels of vibration should be avoided.
See Section
8.2.3 Environmental Specifications
for complete environmental specifications.
CAUTION:
Keep the SECTOR Imager away from direct sources of heat or cold and direct sunlight. Ensure that the rear
cooling vents and the CCD camera tower cooling vents on the SECTOR Imager are
not
blocked.
CAUTION:
Do not place any objects, materials, or liquid containers on top of the SECTOR Imager.
CAUTION:
Falling objects or splashing liquids, including chemically reactive or infectious reagents, can cause damage to
the instrumentation or cause injuries. Avoid handling or storing infectious or radioactive materials near the SECTOR Imager.
2.2
Hazards
This section contains notices and warnings that should be read carefully. Before working with the SECTOR Imager, become familiar
with all safety precautions and regulations concerning the handling of materials and the instrument’s electrical and mechanical
components.
Summary of Contents for SECTOR Image 2400
Page 1: ...SECTOR Imager Models2400 6000 INSTRUMENT MANUAL www mesoscale com ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 ImportantInformation ...
Page 15: ...15 3 SystemDescription ...
Page 28: ...28 4 Installation ...
Page 30: ...30 5 QuickStart ...
Page 36: ...Quick Start 36 Figure 5 8 SECTOR Imager window Run Options dialog box ...
Page 38: ...38 6 UsingtheSECTORImager ...
Page 42: ...42 7 Maintenance ...
Page 45: ...45 8 Appendix ...
Page 54: ...54 9 TechnicalSupport ...
Page 56: ...Technical Support 9 2 ...