37
5.4
Results
The software provides several options for viewing results. Select View
Layer Views from the menu or select a layer from the Layer
menu on the toolbar. Refer to the DISCOVERY WORKBENCH User Guide for details on data viewing options. If you observe the data as
it is acquired in Color Map view, bright wells should display numbers above 2000 and dark wells should display numbers below 200.
shows a Color Map view of the data from the demonstration plate.
Figure 5:9 Demonstration plate results, Color Map layer
When using the SECTOR Imager 6000 demonstration plate sectors 3 and 4 (
) should display four bright wells each, as
. Sectors 1, 2, 5, and 6 should have uniform, low signal levels corresponding to instrument background. The
demonstration plate provided for the 2400/2400A (
) functions similarly, with sectors 5, 12, 14 and 19 displaying bright
wells.
Summary of Contents for SECTOR Image 2400