82
CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
A123
Reference heating element acquisition
chain fault
Contact Technical Service (
see Appendix
)
.
A125
Reference heating element acquisition
chain fault
A 201
Steam generator safety thermostat inter-
vened.
Steam generator or heating element mal-
function.
A 202
heating element safety thermostat inter-
vened.
Steam generator or heating element mal-
function.
A 203
Steam generator malfunction
A 204
heating element malfunction
A 250
Water or condensate in the sterilization
chamber.
do a
RESET
following the instructions.
Thoroughly dry the inside of the sterilization chamber and restart the
cycle.
Do not
insert material impregnated with water or liquids in general into
the chamber.
Drain filter obstructed.
Clean the drain filter
(see the
Appendix
"
MAINTENANCE"
).
air seepage through the gasket.
do a
RESET
following the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with
water.
Restart the cycle.
Vacuum pump failure.
Contact Technical Service
(
see Appendix
).
problem in the hydraulic circuit.
A 251
Water injection pump malfunction.
Contact Technical Service
(
see Appendix
).
problem in the hydraulic circuit.
Steam generator safety thermostat inter-
vened.
Steam generator malfunction