B Classic-28 model
Chamber Dimensions (Depth 450mm / Diameter 280mm / Volume 27.7 litres)
PR
OGR
AM
N
A
ME
CYC
LE
T
YPE
STE
R
ILI
ZATION
TEM
PE
RAT
UR
E
AN
D
TIM
E
DRY
ING
TIM
E
(*
)
TOTAL
CYCL
E
TI
M
E
(A
VE
RA
G
E
LOAD
÷
MA
X
LOAD
)
A
VE
RAGE
H2O
CO
NS
U
M
PTIO
N
(ML
/C
YC
LE
)
A
VE
RAGE
ENER
GY
CO
NS
U
M
PTIO
N
(K
WH
/C
YC
LE
)
MA
XIM
U
M
LOAD
(**
) (
***)
(C1)
UNIVERSA
L
134
PREVACUUM
(3 steps; -0.8
bar each step)
134°C / 273°F
4 minutes
14.5 min
58÷61
859
0,8
5 kg unwrapped metal or polymer instruments and
dental handpieces (max 0.9 kg per tray)
2.5kg single package metal or polymer instruments
and dental handpieces (max 0.9 kg per tray)
1.5kg double package metal or polymer instruments
and dental handpieces (max 0.9 kg per tray)
1 kg unwrapped porous and fabric textiles
0.75 kg single package porous and fabric textiles
0.6kg double package porous and fabric textiles
(C2)
UNIVERSA
L
121
PREVACUUM
(3 steps; -0.8
bar each step)
121°C / 250°F
20 minutes
14.5 min
70÷73
890
0,9
5 kg unwrapped metal or polymer instruments and
dental handpieces (max 0.9kg per tray)
2.5kg single package metal or polymer instruments
and dental handpieces (max 0.9kg per tray)
1.5kg double package metal or polymer instruments
and dental handpieces (max 0.9kg per tray)
1 kg unwrapped porous and fabric textiles
0.75 kg single package porous and fabric textiles
0.6kg double package porous and fabric textiles
(C4) SOLID
134
PREVACUUM
(1 step; -0.8
bar)
134°C / 273°F
4 minutes
14.5 min
43÷46
477
0,7
5 kg unwrapped metal or polymer instruments
(max 0.9kg per tray)
2.5kg single package metal or polymer instruments
(max 0.9kg per tray)
(C5)
HOLLOW
134
PREVACUUM
(3 steps; -0.8
bar each step)
134°C / 273°F
4 minutes
3 min
47÷50
890
0,8
5 kg unwrapped metal or polymer instruments
(max 0.9kg per tray)
HELIX/BD
TEST
PREVACUUM
(3 steps; -0.8
bar each step)
134°C / 273°F
3.5 minutes
1.5 min
44
/
/
Test device only (without another load)
VACUUM
TEST
Vacuum
(-0.8 bar)
/
/
20
/
/
Empty chamber
Vacuum -0.8 bar
Test Time: 27 min
HELIX/BD
TEST (****)
/
/
/
66
/
/
/
(*) Default drying time, but the drying time can be manually increased
(**) Cycles C1 and C
2
are validated to sterilize up to 6 single sided lumened devices no longer than 118 mm and no smaller than 0.8
mm(corresponding to “hollow A” cavities with a length/diameter (L/D) ratio = 1 ≤ L/D ≤ 750). Cycles C4 and C5 are validated to sterilize up
to 6 double sided lumened devices no longer than 115 mm and no smaller than 10 mm .
(***) If the material is sterilized “unwrapped” the sterilized material shall used immediately after sterilization (the cycle shall be considered as
immediate use cycle)
(****) This program enables to execute, in sequence, Vacuum test and Helix/BD test (nevertheless each test is executed separately)
PROGRAM SUMMARy TABLE
120 V~ 60 hz
Summary of Contents for B Classic-28
Page 5: ......
Page 62: ...64 EXAMPLES OF PRINTED REPORTS WITH OPTIONAL PRINTER...