75
ALARM CODES
The
list
of alarm codes and, consequently, the messages displayed on the lCd and relative ReSeT mode, is as follows:
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET PROCEDURE
ERRORS (category E)
E000
blackout
blaCkouT
2
E001
Excessive mains voltage
oVeRVolTaGe
1
E002
Water conductivity threshold 1 exceeded
h2o qualITy InSuFFICIenT
1
E003
Water conductivity threshold 2 exceeded
h2o qualITy InSuFFICIenT
1
E010
door open
dooR open
1
E020
Door lock system activation timeout excee
-
ded (
closing)
dooR loCk TImeouT
1
(then reattempt or turn
off)
E021
Door lock system activation timeout excee
-
ded
(opening)
dooR loCk TImeouT
1
(then reattempt or turn
off)
E030
The water in the feed tank is at minimum
level (mIn)
mIn WaTeR leVel
1
E031
The water in the drain tank is at maximum
level (max)
max dRaIn leVel
1
E040
Tank filling failed (automatic filling)
Top-up pRoblem
1
E041
Too frequent tank filling
(
automatic filling
)
Top-up pRoblem
1
E260
Chamber depressurization too slow
ppd SloW
1
E261
Chamber levelling too slow
leVellInG SloW
1
E900
Vacuum test failed
(
during
TEST PHASE
)
TeST FaIled
2
E901
Vacuum test failed
(
during
STANDBy PHASE
)
TeST FaIled
2
E902
Vacuum test failed
(
vacuum pulse timeout exceeded
)
TeST FaIled
2
E999
manually interrupting the cycle
manual InTeRRupTIon
2
1
= ok (warning)
2
= ok + door unl ReSeT