A
B
A
C
B
10
The
first
and fundamental step in achieving good sterilizer operation, long life and complete use
of its features is a correct, careful installation. moreover, this precaution will avoid the danger of
physical injury or property damage, not to mention malfunctions and damage to the machine. So,
please follow the instructions in this chapter
scrupulously
.
NOTE
CuSTomeR SuppoRT (See appendIx) WIll anSWeR youR queSTIonS
and pRoVIde addITIonal InFoRmaTIon.
The STeRIlIzeR haS paSSed all RequIRed InSpeCTIonS beFoRe
beInG plaCed on The maRkeT. IT doeS noT RequIRe any addITIonal
CalIbRaTIon beFoRe beInG plaCed In SeRVICe.
Dimensions and
weight
a. height (total)
500 mm
b. Width (total)
480 mm
C. Depth (excluding
rear connections)
600 mm
Total weight
60
kg
Electricity.
The electrical system to which the sterilizer will be connected must be suitably dimensioned ba-
sed on the electrical characteristics of the device. This information is shown on the
back of the
machine
.
When installing the sterilizer inside a cabinet, you must
provide adequate space all around the device to pro-
vide effective ventilation as well as
a large enough
opening in the back
that, in addition to allowing the
passage of the power cord will also provide
an ade-
quate air flow
and the consequent
optimum cooling
of the heat exchanger.
The compartment where the steriliser will be kept must
have the following
minimum dimensions
:
Dimensions
a. height
500 mm with FRonT FIllInG kIT
670 mm with FIllInG FRom The Top (door)
b. Width
600 mm
WARNING
CompaRTmenT dImenSIonS leSS Than ThoSe ShoWn may Com-
pRomISe The CoRReCT CIRCulaTIon oF aIR aRound The deVICe
and may noT pRoVIde adequaTe CoolInG, WITh The ConSequenT
deTeRIoRaTIon oF peRFoRmanCe and/oR poSSIble damaGe.
NOTE
IF THE FILLING AND DRAINAGE DOORS ARE NOT ACCESSIBLE FROM THE
TOP AFTER THE DEVICE HAS BEEN BUILT IN, IT IS ADVISABLE TO USE THE
FRONT ATTACHMENTS (FRONT FILLING KIT).
IF The maIn SWITCh IS InaCCeSSIble When InSTalled In The Com-
paRTmenT, uSe an eleCTRIC pluG ThaT InCoRpoRaTeS an on/oFF
SWITCh.
do noT RemoVe The uppeR CoVeR oR any oTheR exTeRnal paRT.
When InSTalled In The CompaRTmenT, The deVICe muST be Com-
pleTe WITh all ITS paRTS.
pleaSe ReFeR To appendIx TeChnICal ChaRaCTeRISTICS FoR Com-
pleTe TeChnICal daTa.
INSTALLATION
INTRODUCTION
COMPARTMENT
DIMENSIONS
FOR BUILT-IN
INSTALLATIONS
Summary of Contents for B Classic-28
Page 5: ......
Page 62: ...64 EXAMPLES OF PRINTED REPORTS WITH OPTIONAL PRINTER...