Water tank cleaning
Boiler gasket replacement
empty the tanks (
C
) containing the autoclave feedwater and wastewater, remove any deposits
around the filter (
D
) on the bottom of the tank (see figure).
After removing and cleaning the filter, wipe the inside of the tanks with a dry cloth and clean
thoroughly.
When cleaning is complete, refit the filter in the tank.
NOTE
do noT uSe CleanInG pRoduCTS InSIde The Tank. uSe a dRy CloTh
only.
It is advisable to have the boiler gasket replaced by an authorised technician, therefore, contact
Technical Service (see
APPENDIX – TECHNICAL SERVICE).
Replace the bacteriological
filter
Dust filter cleaning (ONLY
FOR B VERSIONS)
Remove the dust filter
(A)
from the lower part of the autoclave, thoroughly rinse it with water and
dry it before refitting it.
When filter maintenance is due according to the maintenance schedule or every time you notice
visible clogging of the filter (indicated by the filter markedly turning grey), unscrew the bacterio
-
logical filter
(B)
from its holder and replace it with a new one, screwing it fully down on the union.
NOTE
a SpaRe baCTeRIoloGICal FIlTeR IS pRoVIded WITh The STeRIlIzeR.
To oRdeR FuRTheR FIlTeRS, ReFeR To The appendIx TeChnICal
SeRVICe.