APPENDIX – PROGRAMS
INTRODUCTION
The steam sterilizer is appropriate for almost all materials and instruments, so long as they are able to tolerate, without damage,
a
minimum temperature of 121 °C
(otherwise, you will need to use other low-temperature sterilization systems).
The following material can normally be sterilized with steam:
– Stainless steel surgical/generic instruments;
– Carbon steel surgical/generic instruments;
– Rotating and/or vibrating instruments driven by compressed air (turbines) or mechanical transmission (counter-angles, tooth
scalers);
– mineral-based articles;
– articles made of heat-resistant plastic;
–
Heat-resistant textiles;
– medication materials (gauze, pads, etc.);
– other generic material suitable for autoclave treatment.
NOTE
dependInG on The ConFoRmaTIon oF The maTeRIal (SolId, holloW oR poRouS),
any paCkaGInG (papeR/plaSTIC enVelope, STeRIlIzaTIon papeR, ConTaIneR,
muSlIn napkIn, eTC.) and ITS heaT-ReSISTanCe, IT IS IndISpenSable ThaT you
ChooSe The appRopRIaTe pRoGRam by ReFeRRInG To The Table ShoWn on The
nexT paGe.
WARNING
The deVICe may noT be uSed FoR STeRIlIzaTIon oF FluIdS, lIquIdS oR
phaRmaCeuTICal pRoduCTS.