32
a sterilization cycle consists of a determined number of phases. The number and duration of the
phases can differ for the programs, based on the type of air extraction, sterilization process and
drying method.
The electronic control system monitors the various phases, at the same time checking that the
various parameters are respected; if any type of anomaly is encountered during the cycle, the
program is immediately interrupted, generating an alarm identified by a code, with a relative mes
-
sage explaining the nature of the problem.
With this type of control, when you select a suitable sterilization program, you are guaranteed
perfect sterilization under any conditions.
after inserting the load in the sterilization chamber (taking the precautions described in the sec-
tion
“Preparing the material to be sterilized”
) , select the desired sterilization cycle as follows:
11/04/13 10:54
OK
OK
134°C Universal B
134°C Hollow S
121°C Universal B
OK
134°C Solid S
STERILIZATION
CYCLES
User-defined
Summary of Contents for B Classic-28