WaRnInG meSSaGe
boIleR FIlTeR CleanInG
dooR loCk lubRICaTIon
duST FIlTeR CleanInG
baCTeRIoloGICal FIlTeR
ReplaCemenT
WaTeR Tank CleanInG
boIleR GaSkeT ReplaCemenT
GeneRal SeRVICe
NOTE
To aChIeVe The beST peRFoRmanCe oF The deVICe, ReGulaR maIn-
TenanCe muST be CaRRIed ouT on a ReGulaR baSIS.
peRIodICally a meSSaGe WIll be dISplayed RequeSTInG ThaT The
aboVe maInTenanCe CheCkS be peRFoRmed.
FoR FuRTheR InFoRmaTIon oR In CaSe oF doubT, ConTaCT The
TeChnICal SeRVICe: IF They haVe peRFoRmed ReGulaR maInTe-
nanCe on The deVICe, The TeChnICIan mIGhT haVe alReady CaR-
RIed ouT Some oF TheSe opeRaTIonS (example: baCTeRIoloGICal
FIlTeR oR GaSkeT ReplaCemenT).
keep the following
general warnings
in mind:
–
Do not
wash the sterilizer with direct
jets of water
, either under pressure or sprinkled. Seepage
into electrical and electronic components could damage the functioning of the device or its
internal parts, even irreparably;
–
Do not
use abrasive cloths, metal brushes (or other aggressive materials) or metal-cleaning
products, whether solids or liquids, to clean the device or sterilization chamber;
–
Do not
use chemical products or disinfectants to clean the sterilization chamber. In fact, these
products can damage the sterilization chamber, even irreparably;
–
Do not
allow lime residue or other substances to accumulate in the sterilization chamber or
on the door and its gasket, but periodically remove them. In fact, they can damage these parts
over time in addition to
compromising
the operation of the components installed along the
plumbing circuit.
NOTE
The FoRmaTIon oF WhITe SpoTS on The baSe oF The InTeRnal
WallS oF The STeRIlIzaTIon ChambeR meanS ThaT you aRe uSInG
loW-qualITy demIneRalIzed WaTeR.