36
RUNNING THE CyCLE
press the ok button to start the cycle with the active options selected.
Taking as example the most
complete and
significant
sterilization cycle, i.e. the
134°C UNIVER-
SAL
program, characterised by fractionated pre-vacuum, the cycle sequence is as follows:
WaRmInG up
FIRST VaCuum phaSe
FIRST pReSSuRe RISe
SeCond VaCuum phaSe
SeCond pReSSuRe RISe
ThIRd VaCuum phaSe
ThIRd pReSSuRe RISe
STeRIlIzaTIon
STeam dISChaRGe
dRyInG
VenTIlaTIon
CyCle CompleTIon
If the message
"COMPLETED"
appears, it means that the cycle has completed correctly without
any interruptions for alarms and that
complete asepsis
of the material is guaranteed.
To open the sterilizer door, press the button shown in the figure:
134°C Univ.
4’
134°C
2.12 Bar
Warming up
39:00
134°C Univ.
0’
60°C
1.12 Bar
Completed
00:00
OK
Stampa etichette
N°etich.
Scadenza (gg)
10
20
NO
134°C Univ.
4’
134°C
2.12 Bar
Warming up
39:00
134°C Univ.
0’
60°C
1.12 Bar
Completed
00:00
OK
Stampa etichette
N°etich.
Scadenza (gg)
10
20
NO
RESULT OF
THE CyCLE
DOOR OPENING
AT CyCLE END
134°C Univ.
0’
60°C
1.12 Bar
Completed
00:00