GENERAL PROBLEMS
INTRODUCTION
If you run into a problem or alarm while using the device, you should
not
be immediately concerned.
It may not, in fact, be related to a breakdown but, more probably to an anomalous situation, often merely transitory (such as a
blackout), or incorrect use.
In any case, it is important to first identify the cause of the anomaly and then take suitable corrective action, either autonomously or
with the help of the
Technical Support Department (see Appendix Z)
.
For this purpose, below, we provide instructions for diagnosing and resolving general problems, in addition to a precise description of
the alarm codes, their meaning and their solution.
ANALySIS AND RESOLUTION OF PROBLEMS
If your sterilizer is
not
working correctly, please make the following checks
before
calling the Technical Support department:
PROBLEM
POSSIBLE CAUSE
PROPOSED SOLUTION
The sterilizer does
not
power-on.
The power cord is not plugged-in.
plug it in.
There is no voltage at the socket.
Check the cause of the lack of voltage at and
socket and fix it.
The main switch and/or differential
switch are oFF.
Turn the switch on.
The mains fuses are blown.
Replace with good fuses of equal nominal
value.
(See the
Summary Table
in
Appendix
Technical Characteristics
).
after pressing
START
, the sterilization
cycle does
not
start.
The device is preheating.
Wait for the sterilizer to reach the proper
operating conditions for starting the program.
NOTE:
under normal conditions, the
average preheating time is about
10-15 minutes.
The safety valve has intervened.
locking ring loosened.
presence of anomalous overpressure in
the chamber.
Check that the knurled locking ring is cor-
rectly tightened on the upper part of safety
valve.
DANGER
leT The deVICe
Cool, oR WeaR
GloVeS To aVoId
beInG buRned
When TouChInG
The ValVe.
.
There is water on the support surface of
the sterilizer.
Automatic water filling system tube (op
-
tional) not properly connected.
Check the seal of the unions; if necessary,
refit paying greater attention.
Check that the tubes are completely fitted on
the unions; check that the hose clamps are
fitted.
Steam leaks from the gasket.
at the end of the cycle, clean the gasket and
porthole of the container under pressure.
Check if the gasket is damaged.
Run another cycle and check the situation.