77
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET PROCEDURE
ALARMS (category A)
A250
1st vacuum pulse not reached within the ti-
meout
pV1 TImeouT
2
A251
1st rise back up to atmospheric pressure not
reached within the timeout
aTm1 TImeouT
2
A252
1st pressure pulse not reached within the ti-
meout
pp1 TImeouT
2
A253
2nd vacuum pulse not reached within the ti-
meout
pV2 TImeouT
2
A254
2nd rise back up to atmospheric pressure not
reached within the timeout
aTm2 TImeouT
2
A255
2nd pressure pulse not reached within the
timeout
pp2 TImeouT
2
A256
3rd vacuum pulse not reached within the ti-
meout
pV3 TImeouT
2
A257
3rd rise back up to atmospheric pressure not
reached within the timeout
aTm3 TImeouT
2
A258
3rd pressure pulse not reached within the ti-
meout
ppp TImeouT
2
A260
Chamber depressurization not completed
within the timeout
ppd TImeouT
2
A261
Chamber levelling not completed within the
timeout
leVellInG TImeouT
2
A353
1st drop to atmospheric pressure not comple-
ted within the timeout
dRaIn TImeouT
2
A356
2nd drop to atmospheric pressure not comple-
ted within the timeout
dRaIn TImeouT
2
A360
Vacuum pulsation during drying not executed
within timeout
TIme ouT ppd
2
1
= ok (warning)
2
= ok + door unl ReSeT
Summary of Contents for B Classic-28