75
A
Electromagnetic Compatibilit
y
Guidelines and manufacturer declaration – electromagnetic interference immunity
The customer or user of the MOT-VS700 must ensure that it is operated in suchlike environments.
Interference
immunity tests
IEC 60601 test level
Compliance level
Electromagnetic
environment – guidelines
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV
contact discharge
± 2; 4; 8; 15 kV
air discharge
± 8 kV
contact discharge
± 2; 4; 8; 15 kV
air discharge
Floors should be wood,
concrete or ceramic tile.
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV
for power supply lines
± 1 kV
for input and output lines
± 2 kV
for power supply lines
± 1 kV
for input and output lines
Power supply voltage quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 0,5; 1 kV differential
mode voltage
± 0,5; 1; 2 kV common
mode voltage
± 0,5; 1 kV differential
mode voltage
± 0,5; 1; 2 kV common
mode voltage
Power supply voltage quality
should be that of a typical
commercial or hospital
environment.
Voltage drops, short
interruptions and
supply voltage
IEC 61000-4-11
0 % U
T
for 0.5 cycles
0 % U
T
for 1 cycles
70 % U
T
for 25 / 30 cycles
0 % U
T
for 250 / 300 cycles
0 % U
T
for 0.5 cycles
0 % U
T
for 1 cycles
70 % U
T
for 25 / 30 cycles
0 % U
T
for 250 / 300 cycles
Power supply voltage quality
should be that of a typical
commercial or hospital
environment.
If the user of the
MOT-VS700 need to
continue operation during
a mains power interruption,
it is recommended that the
MOT-VS700 be powered
by an uninterruptible power
supply or battery.
Power frequency
(50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
characteristic of a typical
location in a typical
commercial or hospital
environment.
Info: U
T
is the AC voltage before application of the test level